产品搜索:
产品资料
产品[

OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12

]资料
如果您对该产品感兴趣的话,可以 sendmsg
产品名称: OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12
产品型号: OM-CP-HITEMP140
产品厂商: 美国Omega热电偶
产品文档: 无相关文档
简单介绍
    OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12由美国omega原厂渠道的深圳鑫博恒业科技有限公司特价提供,OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12精度高,寿命长,OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12是值得信赖的记录仪.
OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12 的详细介绍

OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12

OM-CP-HITEMP140-M12

数据采集 » » OM-CP-HITEMP140-M12 

高温度数据记录仪内置M12连接器
M12连接器–探针快速交换RTD探头以满足您的应用程序
测量范围为200~850°
(328至1562°)
(0.05°)(0.09°)准确度
用户可更换电池
触发设置
可编程启动和停止时间
电池寿命指示器

产品描述
The OM-CP-HITEMP140-M12 is a rugged and versatile high temperature data logger featuring an M12 probe connector. Compatible with a wide variety of M12 RTD probe styles, this data logger is capable of measuring up to 850°C (1562°F) (probe dependent). With the ability to change probes as needed, this device satisfies a multitude of application needs with one powerful data logger versus the need for multiple loggers.

The OM-CP-HITEMP140 data logger has a water tight body enclosure made of food grade stainless steel. It can be placed in environments with temperatures as high as 140°C (284°F) making it suitable for a wide range of applications including autoclave verification, food preparation and processing, environmental studies, well monitoring, washer disinfectors and pasteurization. The OM-CP-HITEMP140-M12 records and stores up to 43,690 time stamped readings and is equipped with non-volatile solid state memory which retains data even if the battery becomes discharged.

Using the OM-CP-HITEMP140-M12 software, starting, stopping and downloading the OM-CP-HITEMP140-M12 is simple and easy. Graphical, tabular and summary data is provided for analysis and data can be viewed in °C, °F, K or °R. The data can also be automatically exported to Excel® for further calculations.

The OM-CP-MULTIMOUNT-Z is a versatile mount or stand for use with the OM-CP-HITEMP140 series of data loggers. It can be used to stabilize a logger inside an autoclave, or screwed to a flat surface to create an anchored base. The OM-CP-MULTIMOUNT-Z is made of stainless steel and is able to withstand temperatures up to 150°C (302°F) making it ideal for use in autoclave sterilization processes.

The OM-CP-ALUCASE-14 and OM-CP-ALUCASE-18 are high quality, multi-purpose briefcases, designed for the protection, transport and storage of data loggers, equipment and accessories. These durable aluminum cases include pick and pluck foam inserts, allowing the user to customize a snug fit and secure any combination of products. The briefcase frame is strong yet surprisingly light weight with a comfortable ergonomically upholstered handle. Two sizes are available: a compact 36 cm (14") width (OM-CP-ALUCASE-14) and large 47 cm (18") width (OM-CP-ALUCASE-18), both featuring an egg crate foam lined lid and keyed locks.

The OM-CP-IFC406 is a multiplexer data logger interface for compatible OM-CP Series data loggers. The OM-CP-IFC406 allows up to six data loggers to be connected into one interface. Up to three OM-CP-IFC406 units may be daisy-chained together to communicate with a total of 18 data loggers through one USB port. To connect multiple OM-CP-IFC406 multiplexer interfaces together, simply join the units side by side, making sure the spring pin contacts are connected and magnetically joined.

Specifications
Temperature Sensor: Pt100 RTD probe with M12 connector (probe sold separately)
Measurement Range: 18 to 400 Ω; -200 to 850°C (-328 to 1562°F) (probe dependent)
Temperature Resolution: 0.0001 Ω; 0.01°C (0.018°F) (probe dependent)
Calibrated Accuracy: ±0.015 Ω; ±0.05°C (±0.09°F) (probe dependent)
Calibrated Accuracy Range: 18 to 200 Ω; -200 to 265°C (-328 to 509°F)
Connector Type: M12 female, 4-pin
Reading Rate: 4 readings per second up to 1 reading every 24 hours
Memory: 43,690 readings
Start Modes:
• Software programmable immediate start
• Delay start up to 18 months in advance
Stop Modes: Manual or timed (specific date and time)
Real Time Recording: May be used with PC to monitor and record data in real time
Password Protection: An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password
Trigger Settings: High and low limits may be set. Once data meets or exceed sets limits, the device will record to memory. Bi-level start and stop triggers can also be programmed. Users can specify the number of readings to take after the device triggers
Readings in Trigger Settings Mode: 18,724 readings
Memory Wrap Around: Yes
Battery Type: 3.6V high-temperature lithium battery included; user replaceable
Battery Life: 2 years typical [1 minute reading rate at 25°C (77°F)]
Calibration: Digital calibration through software
Calibration Date: Automatically recorded within device
Data Format: Date and time stamped °C, °F, °R, K
Time Accuracy: ±1 minute/month at 25°C (77°F)
Computer Interface: OM-CP-IFC400 or OM-CP-IFC406 USB docking station required; 125,000 baud
Operating Environment: -40 to 140°C (-40 to 284°F), 0 to 100% RH non-condensing, 0.002 to 60 psia
Software: XP SP3/Vista/7 and 8 (32-bit and 64-bit)
IP Rating: IP40-logger alone (with no probe connected); the logger will inherit IP rating of the attached probe (up to, and including IP68/IP69K)
Dimensions (Body): 54.5 H x 24.6 mm dia (2.15 x 0.97")
Weight: 85 g (3.0 oz)
Materials (Body): 316 Stainless Steel

产品编号 产品描述
  
OM-CP-HITEMP140-M12 High temperature data logger with M12 connector

OM-CP-HITEMP140-M12-CERT High temperature data logger with M12 connector and NIST calibration certificate

Accessories
  
OM-CP-IFC400 Docking station (for single data logger) with USB cable, software and manual

OM-CP-IFC406 Multiplexer data logger interface (accepts up to 6 data loggers) with USB cable, software and manual

OM-CP-BAT110 Replacement 3.6 V lithium battery

OM-CP-SVP-SYSTEM FDA 21 CFR part 11 compliant IQ/OQ/PQ secure software validation workbook and software package (unlimited users, license per computer)
 
OM-CP-MULTIMOUNT-Z Mount/stand for OM-CP-HITEMP140 data loggers
 
OM-CP-ALUCASE-14 Aluminum data logger briefcase, 36.8 x 29.2 cm (14.5 x 11.5")
 
OM-CP-ALUCASE-18 Aluminum data logger briefcase, 47 x 38.1 cm (18.5 x 15")

 OM-CP-HITEMP140-M12

OM-CP-HITEMP140-M12记录仪 美国omega
OM-CP-HITEMP140-M12-CERT记录仪 美国omega


OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12衍生阅读

OM-CP-HITEMP140-M12美国OMEGA OM-CP-HITEMP140-M12相关产品

产品留言
标题
联系人
联系电话
内容
验证码
点击换一张
注:1.可以使用快捷键Alt+S或Ctrl+Enter发送信息!
2.如有必要,请您留下您的详细联系方式!


img15
img16
深圳市鑫博恒业科技有限公司 版权所有
电话:0755-82768776 82736276    传真:0755-83041664  电邮 :info@xb5j.cn
地址:深圳市福田区华强北路华强广场A栋17A(华强北地铁站-A口)  邮编:518031
粤ICP备09037369号  

著作权声明
  

粤公网安备 44030402001378号